The only medical gas outlet that accepts both Chemetron and Ohmeda adapters (patent #5,562,121).
Push-button mechanism for quick release of Chemetron adapter; Ohmeda adapter has a twist mechanism for quick release.
Chemetron adapter can support secondary equipment, including a suction regulator with a half-gallon collection bottle.
Eliminates the need for hospital personnel to carry both styles of adapters to ensure proper connection.
Fail-safe color coding and safety keyed to ensure appropriate gas.
Valve body O-ring easily accessible for maintenance; inlet tube has secondary check unit to permit removal of valve body without shutting off gas to the outlet.
Cleaned for oxygen use.
UL and CSA* listed. NFPA compliant.
The Allied Connect2® medical gas outlet stations provide convenient access to medical gases in patient care locations. They accept either a Chemetron or Ohmeda adapter in the same outlet.
The type K copper inlet tube has a 1/2" OD connection, and can rotate 360 degrees to accommodate any field piping arrangement. The primary gas check can be serviced without shutting off the gas supply.
Elements Connect2® Quick-Connect Medical Gas Outlets The outlets are field-gangable in any order desired by the facility. Outlets are spaced 5" (12.7 cm) center to center. Outlet finish plates and rough-in plates are pin indexed using steel pins for positive cross-connection prevention. Primary valves are separate from the finish plate, allowing easy testing for leaks and cross connection. Primary valves are made of brass and contain a chromeplated self-sealing poppet. Primary valves are field-adjustable to compensate for variations in plaster thickness of 1/2" to 3/4". Outlets may be special ordered with extended valve bodies that allow for an additional 3/8" or 3/4" adjustment.>p> Available for : Oxygen, Air, Vaccum, Nitrous Oxide and Carbon Dioxide