EMT MTV-100 Resuscitator with 6' O2 hose

SKU
L576-010
$394.54
-
+
Overview
Most advanced design in 02 powered manually triggered breathing devices. Limits gas flow to 40 LPM during manually triggered breath, however control circuit will allow up to 115 LPM during a spontaneous breath. Material: Body-Liquid Crystal Polymer Cap- Liquid Crystal Polymer; Button- Liquid Crystal Polymer; Outlet- Polysulfone; Inlet Fitting- Plated Brass. Standard male oxygen DISS(swivel) The EMT Resuscitator may be attached to a portable pressure regulated oxygen cylinder or a central oxygen source. It may be used with a mask, endotracheal tube or tracheotomy tube. Advanced pneumatic technology limits and regulates gas flow to 40 lpm during the manually triggered breath, but will allow flow rates of up to 115 lpm during a spontaneous breath. Addressing AHA guidelines, the EMT Resuscitator delivers a constant preset flow of 40 lpm, reducing the possibility of gastric insufflation during mask ventilation and improving the distribution of gas throughout the lung when ventilating through an endotracheal or tracheotomy tube. The EMT Resuscitator features built in antiasphyxiation ports that allow the patient to breathe outside mbient air in the event of source gas failure. In addition, a 360° swivel oxygen inlet attachment eliminates the problem of line twisting and tangling. Operating Temperature: -10oF to 125oF Filter: 25 micron, wire mesh Outlet: 22mm outside diameter x 15mm inside diameter Supply Pressure: 50+,-5 psig Pressure to initiate a breath: -1 to -2.5cmH2O Manually Triggered Flow: 40+,-5 LPM @50 psi supply Demand Flow: 100LPM at 0 to -5cm H2O pressure Pressure Limit: 60 +- 5cm H2O Secondary Pressure Relief: 75 +- 5 cm H2O pressure Expiratory Resistance: Less than 2cm H2O @ 100LPM flow Work of Breathing: -5cm H2O pressure @ 115 LPM flow Anti-Asphyxiation Entrainment: 55 liters per minute minimum @ -5cm H2O pressure

Most advanced design in 02 powered manually triggered breathing devices. Limits gas flow to 40 LPM during manually triggered breath, however control circuit will allow up to 115 LPM during a spontaneous breath.

Material: Body-Liquid Crystal Polymer Cap- Liquid Crystal Polymer; Button- Liquid Crystal Polymer; Outlet- Polysulfone; Inlet Fitting- Plated Brass. Standard male oxygen DISS(swivel) The EMT Resuscitator may be attached to a portable pressure regulated oxygen cylinder or a central oxygen source. It may be used with a mask, endotracheal tube or tracheotomy tube.

Advanced pneumatic technology limits and regulates gas flow to 40 lpm during the manually triggered breath, but will allow flow rates of up to 115 lpm during a spontaneous breath. Addressing AHA guidelines, the EMT Resuscitator delivers a constant preset flow of 40 lpm, reducing the possibility of gastric insufflation during mask ventilation and improving the distribution of gas throughout the lung when ventilating through an endotracheal or tracheotomy tube. The EMT Resuscitator features built in antiasphyxiation ports that allow the patient to breathe outside mbient air in the event of source gas failure. In addition, a 360° swivel oxygen inlet attachment eliminates the problem of line twisting and tangling.

Operating Temperature: -10oF to 125oF

Filter: 25 micron, wire mesh

Outlet: 22mm outside diameter x 15mm inside diameter

Supply Pressure: 50+,-5 psig

Pressure to initiate a breath: -1 to -2.5cmH2O

Manually Triggered Flow: 40+,-5 LPM @50 psi supply

Demand Flow: 100LPM at 0 to -5cm H2O pressure

Pressure Limit: 60 +- 5cm H2O

Secondary Pressure Relief: 75 +- 5 cm H2O pressure Expiratory

Resistance: Less than 2cm H2O @ 100LPM flow

Work of Breathing: -5cm H2O pressure @ 115 LPM flow

Anti-Asphyxiation Entrainment: 55 liters per minute minimum @ -5cm H2O pressure

 

Write Your Own Review
You're reviewing:EMT MTV-100 Resuscitator with 6' O2 hose