<p><span style="display: inline !important; float: none; background-color: transparent; color: rgb(51, 51, 51); font-family: sans-serif,Arial,Verdana," trebuchet="">Vacutron Intermittent model manufactured by Allied Healthcare Products - Chemetron Corporation</span></p>
<p>Intermittent suction regulators are designed to deliver intermittent vacuum only. This regulator is used to ensure that constant suction is never delivered.</p>
<p style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet=""><span style="display: inline !important; float: none; background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet="">Configurations:</span></p>
<p style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet=""><strong style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet="">607-22-16-1108 </strong></p>
<ul style="background-color: transparent; color: rgb(51, 51, 51); font-family: sans-serif,Arial,Verdana," trebuchet="">
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet="">Outlet to the patient fitting: diss male</li>
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet="">Inlet to the wall fitting: Chemetron</li>
</ul>
<p style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet=""><b style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet="">607-22-16-1104 </b></p>
<ul style="background-color: transparent; color: rgb(51, 51, 51); font-family: sans-serif,Arial,Verdana," trebuchet="">
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text;" trebuchet="">Outlet to the patient fitting: diss male</li>
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text;" trebuchet="">Inlet to the wall fitting: Oxequip Med Star</li>
</ul>
<p style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet=""><b style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet="">607-22-16-1106</b></p>
<ul style="background-color: transparent; color: rgb(51, 51, 51); font-family: sans-serif,Arial,Verdana," trebuchet="">
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text;" trebuchet="">Outlet to the patient fitting: diss male</li>
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text;" trebuchet="">Inlet to the wall fitting: Ohmeda</li>
</ul>
<p style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet=""><b style="background-color: transparent; color: rgb(51, 51, 51); cursor: text; font-family: sans-serif,Arial,Verdana," trebuchet="">607-22-16-1105</b></p>
<ul style="background-color: transparent; color: rgb(51, 51, 51); font-family: sans-serif,Arial,Verdana," trebuchet="">
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text;" trebuchet="">Outlet to the patient fitting: diss male</li>
<li style="background-color: transparent; color: rgb(51, 51, 51); cursor: text;" trebuchet="">Inlet to the wall fitting: Puritan bennett</li>
</ul>
Specifications: <br>
Operating Modes: OFF and Intermittent<br>
Regulated Vacuum Range: 0-150 mm Hg<br>>
Cycle Times: 17 seconds of suction followed by 10 seconds without suction<br>
Safety Relief Valves: 1 @ 170 mm Hg<br>
Vacuum Gauge Scale: 0-300 mm Hg in 5 mm Hg increments<br>
0-40 kPa in 2 kPa increments<br>
Dimensions: 6.8"H x 2.65" W x 4.07"<br>
D (174 mm x 67 mm x 104 mm)<br>
Warranty: 10 years